Severity
Moderate
FDA Devices recall · Reported January 21, 2026
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may no…
Canon Medical System, USA, INC. recalled VANTAGE TITAN 3T Model MRT-3010/MEXL-3010 - a moderate-severity action.
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010 was recalled by Canon Medical System, USA, INC. in January 21, 2026. Reason: There is a potential for formation of ice occurring in the venting system of the superconducting magnet of th…. Check the official notice for the remedy. Verify recall #Z-1018-2026 with the FDA Devices before acting.
The recall
Canon Medical System, USA, INC. issued this moderate-severity FDA Devices recall-There is a potential for formation of ice occurring in the venting system of the superconducting magnet of th….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1018-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1018-2026) was formally reported on January 21, 2026, with the manufacturer initiating the action on December 19, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Canon Medical System, USA, INC. is listed as the recalling firm, operating out of Tustin, CA. Federal records list the affected scope as 22 systems.
The documented reason for this recall is: There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed ven… Distribution data in the federal record shows the product reached: U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
22 systems
Related Recalls
6
3 from same agency
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1018-2026 |
| Date reported | January 21, 2026 |
| Date initiated | December 19, 2025 |
| Recalling firm | Canon Medical System, USA, INC. |
| Firm location | Tustin, CA |
| Affected scope | 22 systems |
| Distribution | U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 21, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.