FDA Devices recall · Reported February 26, 2014
TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM O TRILOGY ACET SHELL 50MM O TRILOGY ACET SHELL 52MM O TRILOGY ACET SHELL 54MM O TRILOGY ACET SHELL 56MM O TRILOGY ACET SHELL 58MM O TRILOGY ACET SHELL 60MM O TRILOGY ACET SHELL 62MM O TRILOGY ACET SHELL 64MM O TRILOGY ACET SHELL 66MM O TRILOGY ACET SHELL 68MM O TRILOGY ACET SHELL 70MM O TRILOGY ACET S
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was deter…
Zimmer, Inc. recalled TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY … - a moderate-severity action.
TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY … was recalled by Zimmer, Inc. in February 26, 2014. Reason: Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZM…. Check the official notice for the remedy. Verify recall #Z-1024-2014 with the FDA Devices before acting.
The recall
Zimmer, Inc. issued this moderate-severity FDA Devices recall-Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZM….
- Moderate
- severity level
- Class II
- classification
- February 26, 2014
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1024-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM O TRILOGY ACET SHELL 50MM O TRILOGY ACET SHELL 52MM O TRILOGY ACET SHELL 54MM O TRILOGY ACET SHELL 56MM O TRILOGY ACET SHELL 58MM O TRILOGY ACET SHELL 60MM O TRILOGY ACET SHELL 62MM O TRILOGY ACET SHELL 64MM O TRILOGY ACET SHELL 66MM O TRILOGY ACET SHELL 68MM O TRILOGY ACET SHELL 70MM O TRILOGY ACET SHELL 72MM O TRILOGY ACET SHELL 74MM O TRILOGY ACET SHELL 76MM O TRILOGY ACET SHELL 78MM O TRILOGY ACET SHELL 80MM O BONE SCREW 4.5X15 SELF-TA BONE SCREW 4.5X20 SELF-TA BONE SCREW 4.5X25 SELF-TA BONE SCREW 4.5X30 SELF-TA BONE SCREW 4.5X35 SELF-TA BONE SCREW 4.5X40 SELF-TA BONE SCREW 4.5X50 SELF-TA BONE SCREW 4.5X60 SELF-TA BONE SCREW 6.5X15 SELF-TA BONE SCREW 6.5X20 SELF-TA BONE SCREW 6.5X25 SELF-TA BONE SCREW 6.5X30 SELF-TA BONE SCREW 6.5X35 SELF-TA BONE SCREW 6.5X40 SELF-TA BONE SCREW 6.5X50 SELF-TA BONE SCREW 6.5X60 SELF-TA BONE SCREW 6.5X70 SELF-TA BONE SCREW 6.5X80 SELF-TA
Reason for recall
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1024-2014 |
| Date reported | February 26, 2014 |
| Date initiated | October 11, 2013 |
| Recalling firm | Zimmer, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 7,044,680 total |
| Distribution | Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portu… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1024-2014) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls