Severity
Moderate
FDA Devices recall · Reported February 26, 2014
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was deter…
Zimmer, Inc. recalled TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY … - a moderate-severity action.
TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY … was recalled by Zimmer, Inc. in February 26, 2014. Reason: Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZM…. Check the official notice for the remedy. Verify recall #Z-1024-2014 with the FDA Devices before acting.
The recall
Zimmer, Inc. issued this moderate-severity FDA Devices recall-Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZM….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1024-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1024-2014) was formally reported on February 26, 2014, with the manufacturer initiating the action on October 11, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Zimmer, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records list the affected scope as 7,044,680 total.
The documented reason for this recall is: Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations cond… Distribution data in the federal record shows the product reached: Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Ze…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
7,044,680 total
Related Recalls
6
3 from same agency
TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM O TRILOGY ACET SHELL 50MM O TRILOGY ACET SHELL 52MM O TRILOGY ACET SHELL 54MM O TRILOGY ACET SHELL 56MM O TRILOGY ACET SHELL 58MM O TRILOGY ACET SHELL 60MM O TRILOGY ACET SHELL 62MM O TRILOGY ACET SHELL 64MM O TRILOGY ACET SHELL 66MM O TRILOGY ACET SHELL 68MM O TRILOGY ACET SHELL 70MM O TRILOGY ACET SHELL 72MM O TRILOGY ACET SHELL 74MM O TRILOGY ACET SHELL 76MM O TRILOGY ACET SHELL 78MM O TRILOGY ACET SHELL 80MM O BONE SCREW 4.5X15 SELF-TA BONE SCREW 4.5X20 SELF-TA BONE SCREW 4.5X25 SELF-TA BONE SCREW 4.5X30 SELF-TA BONE SCREW 4.5X35 SELF-TA BONE SCREW 4.5X40 SELF-TA BONE SCREW 4.5X50 SELF-TA BONE SCREW 4.5X60 SELF-TA BONE SCREW 6.5X15 SELF-TA BONE SCREW 6.5X20 SELF-TA BONE SCREW 6.5X25 SELF-TA BONE SCREW 6.5X30 SELF-TA BONE SCREW 6.5X35 SELF-TA BONE SCREW 6.5X40 SELF-TA BONE SCREW 6.5X50 SELF-TA BONE SCREW 6.5X60 SELF-TA BONE SCREW 6.5X70 SELF-TA BONE SCREW 6.5X80 SELF-TA
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1024-2014 |
| Date reported | February 26, 2014 |
| Date initiated | October 11, 2013 |
| Recalling firm | Zimmer, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 7,044,680 total |
| Distribution | Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portu… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 26, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.