FDA Devices recall · Reported February 26, 2014
12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM LD/FX SZ 14X13 VERSYS CEM LD/FX SZ 16X14 VERSYS CEM LD/FX ST 18X15 VERSYS PRESS FIT LD/FX ST VERSYS PRESS FIT LD/FX SZ VERSYS ADVOCATE CEMENTED VERSYS ADVOCATE V-LIGN CE VERSYS CEMENTED FEM STEM VERSYS CT FEM STEM 11X120 VERSYS CT FEM STEM 12X125 VERSYS CT FEM STEM 13X130 VERSYS CT FEM STEM 14X135 VERSYS CT FEM STEM 15X140 VERSYS CT FEM STEM 16X145 VERSYS CT FEM STEM 17X150 VERSYS CEM/REV
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was deter…
Zimmer, Inc. recalled 12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS C… - a moderate-severity action.
12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS C… was recalled by Zimmer, Inc. in February 26, 2014. Reason: Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZM…. Check the official notice for the remedy. Verify recall #Z-1029-2014 with the FDA Devices before acting.
The recall
Zimmer, Inc. issued this moderate-severity FDA Devices recall-Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZM….
- Moderate
- severity level
- Class II
- classification
- February 26, 2014
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1029-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
12/14 UNIPOLAR FEMORAL HE VERSYS CEM LD/FX SZ 11X12 VERSYS CEM LD/FX SZ 12X12 VERSYS CEM LD/FX SZ 14X13 VERSYS CEM LD/FX SZ 16X14 VERSYS CEM LD/FX ST 18X15 VERSYS PRESS FIT LD/FX ST VERSYS PRESS FIT LD/FX SZ VERSYS ADVOCATE CEMENTED VERSYS ADVOCATE V-LIGN CE VERSYS CEMENTED FEM STEM VERSYS CT FEM STEM 11X120 VERSYS CT FEM STEM 12X125 VERSYS CT FEM STEM 13X130 VERSYS CT FEM STEM 14X135 VERSYS CT FEM STEM 15X140 VERSYS CT FEM STEM 16X145 VERSYS CT FEM STEM 17X150 VERSYS CEM/REV/CALCAR 13X VERSYS CEM/REV/CALCAR 15X VERSYS CEM/REV/CALCAR 17X VERSYS CEM/REV/CALCAR 19X
Reason for recall
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1029-2014 |
| Date reported | February 26, 2014 |
| Date initiated | October 11, 2013 |
| Recalling firm | Zimmer, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 7,044,680 total |
| Distribution | Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portu… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1029-2014) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026