PlainRecalls
FDA Devices Moderate Class II Ongoing

HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 20mmx10mm UPN: M00553560

Reported: February 5, 2025 Initiated: December 20, 2024 #Z-1036-2025

Product Description

HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 20mmx10mm UPN: M00553560

Reason for Recall

Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.

Details

Units Affected
1102 units (OUS)
Distribution
Nationwide including Puerto Rico Foreign: To be provided
Location
Marlborough, MA

Frequently Asked Questions

What product was recalled?
HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 20mmx10mm UPN: M00553560. Recalled by Boston Scientific Corporation. Units affected: 1102 units (OUS).
Why was this product recalled?
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 5, 2025. Severity: Moderate. Recall number: Z-1036-2025.