Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.

Reported: January 14, 2026 Initiated: December 12, 2025 #Z-1036-2026 1,301 units units

Integra LifeSciences Corp. (NeuroSciences) issued this FDA Devices recall on January 14, 2026. Classified as Moderate severity (Class II). Approximately 1,301 units units are affected. The recall was issued because: Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-1036-2026) was formally reported on January 14, 2026, with the manufacturer initiating the action on December 12, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Integra LifeSciences Corp. (NeuroSciences) is listed as the recalling firm, operating out of Princeton, NJ. Federal records indicate 1,301 units units are affected.

The documented reason for this recall is: Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits. Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. Because this recall is recent, remedy windows and replacement inventory are most likely still actively available from the firm or retailer. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

1,301 units

Related Recalls

6

6 from same agency

Product Description

Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.

Reason for Recall

Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.

Details

Units Affected
1,301 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
Location
Princeton, NJ

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1036-2026
Date reported January 14, 2026
Date initiated December 12, 2025
Recalling firm Integra LifeSciences Corp. (NeuroSciences)
Units affected 1,301 units
Distribution Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countr…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1,301 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.. Recalled by Integra LifeSciences Corp. (NeuroSciences). Units affected: 1,301 units.
Why was this product recalled?
Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 14, 2026. Severity: Moderate. Recall number: Z-1036-2026.
Where was the recalled product distributed?
Distribution: Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the)..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1036-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).