Digital RID Plate Reader and Software Product Code: AD400

Reported: January 25, 2017 Initiated: October 9, 2012 #Z-1055-2017

Product Description

Digital RID Plate Reader and Software Product Code: AD400

Reason for Recall

If a control ring is marked after reading, the software will not flag results that are out of the specified QC range.

Details

Distribution
U.S. distribution to the following; MA, GA, NY, IA, CA, IL, UT, NJ, NC, MD, MN. Foreign distribution to the following; UK, Germany, Spain, France.
Location
Birmingham

Frequently Asked Questions

What product was recalled?
Digital RID Plate Reader and Software Product Code: AD400. Recalled by The Binding Site Group, Ltd..
Why was this product recalled?
If a control ring is marked after reading, the software will not flag results that are out of the specified QC range.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 25, 2017. Severity: Moderate. Recall number: Z-1055-2017.