Synthes Driving Cap/Threaded Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their function is to insert a nail. The Synthes Driving Cap/Threaded is threaded into the insertion handle and secured with a wrench. If desired, the nail can then be inserted using light hammer blows.

Reported: February 26, 2014 Initiated: January 7, 2014 #Z-1057-2014

Product Description

Synthes Driving Cap/Threaded Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their function is to insert a nail. The Synthes Driving Cap/Threaded is threaded into the insertion handle and secured with a wrench. If desired, the nail can then be inserted using light hammer blows.

Reason for Recall

The Synthes Driving Cap/Threaded can bind in the IM Nail Radiolucent Insertion Handles when assembled becoming difficult to separate.

Details

Recalling Firm
Synthes, Inc.
Units Affected
143
Distribution
USA (nationwide) Distribution.
Location
West Chester, PA

Frequently Asked Questions

What product was recalled?
Synthes Driving Cap/Threaded Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their function is to insert a nail. The Synthes Driving Cap/Threaded is threaded into the insertion handle and secured with a wrench. If desired, the nail can then be inserted using light hammer blows.. Recalled by Synthes, Inc.. Units affected: 143.
Why was this product recalled?
The Synthes Driving Cap/Threaded can bind in the IM Nail Radiolucent Insertion Handles when assembled becoming difficult to separate.
Which agency issued this recall?