FDA Devices recall · Reported February 26, 2014

Synergy XVI XVI can incorrectly calculate the target position of the treatment table, because of an error that does not occur very frequently. This is the sequence of events when this error occurs - (1) You do an online VolumeView acquisition and registration. (2) You accept the registration and the Table Move Assistant dialog box appears. (3) The Relative Actual column of the Table Move Assistant dialog box incorrectly appears blank. (4) When this occurs, XVI does not read the Absolute Actua

XVI can incorrectly calculate the target position of the treatment table.

Recall #
Z-1059-2014
Affected scope
1028
Initiated
November 25, 2013
Verify with FDA Devices →
View my saved recalls

Elekta, Inc. recalled Synergy XVI XVI can incorrectly calculate the target position of the treatment table, … - a moderate-severity action.

Synergy XVI XVI can incorrectly calculate the target position of the treatment table, … was recalled by Elekta, Inc. in February 26, 2014. Reason: XVI can incorrectly calculate the target position of the treatment table.. Check the official notice for the remedy. Verify recall #Z-1059-2014 with the FDA Devices before acting.

The recall

Elekta, Inc. issued this moderate-severity FDA Devices recall-XVI can incorrectly calculate the target position of the treatment table..

Moderate
severity level
1K units
affected scope
Class II
classification
February 26, 2014
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1059-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Synergy XVI XVI can incorrectly calculate the target position of the treatment table, because of an error that does not occur very frequently. This is the sequence of events when this error occurs - (1) You do an online VolumeView acquisition and registration. (2) You accept the registration and the Table Move Assistant dialog box appears. (3) The Relative Actual column of the Table Move Assistant dialog box incorrectly appears blank. (4) When this occurs, XVI does not read the Absolute Actual position of the treatment table. This causes XVI to calculate the treatment table moves from the zero position of the treatment table (+0.0, + 0.0, + 0.0). (5) In XVI R5.0, if the Relative Set values are less than or equal to + or -2.5cm (for earlier releases of XVI, the value is + or -2cm), you can do remote automatic table movement (RATM) from the function keypad (FKP). (6) When you press the Table ASU (automatic setup) buttons, the treatment table moves to a position that was incorrectly calculated from XVI.

Reason for recall

XVI can incorrectly calculate the target position of the treatment table.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1059-2014
Date reported February 26, 2014
Date initiated November 25, 2013
Recalling firm Elekta, Inc.
Firm location Atlanta, GA
Affected scope 1028
Distribution Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL, AK, AZ, AR, CA, CT, FL, GA, ID, IL, KY, LA, MD, MA, MI, MN, MO, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, VT, VA, WA, WV, WI, WY and DC., …

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1028 units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1059-2014) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Synergy XVI XVI can incorrectly calculate the target position of the treatment table, because of an error that does not occur very frequently. This is the sequence of events when this error occurs - (1) You do an online VolumeView acquisition and registration. (2) You accept the registration and the Table Move Assistant dialog box appears. (3) The Relative Actual column of the Table Move Assistant dialog box incorrectly appears blank. (4) When this occurs, XVI does not read the Absolute Actual position of the treatment table. This causes XVI to calculate the treatment table moves from the zero position of the treatment table (+0.0, + 0.0, + 0.0). (5) In XVI R5.0, if the Relative Set values are less than or equal to + or -2.5cm (for earlier releases of XVI, the value is + or -2cm), you can do remote automatic table movement (RATM) from the function keypad (FKP). (6) When you press the Table ASU (automatic setup) buttons, the treatment table moves to a position that was incorrectly calculated from XVI.. Recalled by Elekta, Inc.. Units affected: 1028.
Why was this product recalled?
XVI can incorrectly calculate the target position of the treatment table.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 26, 2014. Severity: Moderate. Recall number: Z-1059-2014.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL, AK, AZ, AR, CA, CT, FL, GA, ID, IL, KY, LA, MD, MA, MI, MN, MO, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, VT, VA, WA, WV, WI, WY and DC., and the countries of Algeria, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom and Venezuela.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported February 26, 2014. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026