Synergy XVI XVI can incorrectly calculate the target position of the treatment table, because of an error that does not occur very frequently. This is the sequence of events when this error occurs - (1) You do an online VolumeView acquisition and registration. (2) You accept the registration and the Table Move Assistant dialog box appears. (3) The Relative Actual column of the Table Move Assistant dialog box incorrectly appears blank. (4) When this occurs, XVI does not read the Absolute Actua
Product Description
Synergy XVI XVI can incorrectly calculate the target position of the treatment table, because of an error that does not occur very frequently. This is the sequence of events when this error occurs - (1) You do an online VolumeView acquisition and registration. (2) You accept the registration and the Table Move Assistant dialog box appears. (3) The Relative Actual column of the Table Move Assistant dialog box incorrectly appears blank. (4) When this occurs, XVI does not read the Absolute Actual position of the treatment table. This causes XVI to calculate the treatment table moves from the zero position of the treatment table (+0.0, + 0.0, + 0.0). (5) In XVI R5.0, if the Relative Set values are less than or equal to + or -2.5cm (for earlier releases of XVI, the value is + or -2cm), you can do remote automatic table movement (RATM) from the function keypad (FKP). (6) When you press the Table ASU (automatic setup) buttons, the treatment table moves to a position that was incorrectly calculated from XVI.
Reason for Recall
XVI can incorrectly calculate the target position of the treatment table.
Details
- Recalling Firm
- Elekta, Inc.
- Units Affected
- 1028
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL, AK, AZ, AR, CA, CT, FL, GA, ID, IL, KY, LA, MD, MA, MI, MN, MO, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, VT, VA, WA, WV, WI, WY and DC., and the countries of Algeria, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom and Venezuela
- Location
- Atlanta, GA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11