Human lgG4 Subclass Liquid Reagent Kits For use on the Roche Cobas"6000 This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents Product Code LK009.CB

Reported: February 1, 2017 Initiated: September 11, 2014 #Z-1069-2017

Product Description

Human lgG4 Subclass Liquid Reagent Kits For use on the Roche Cobas"6000 This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents Product Code LK009.CB

Reason for Recall

Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

Details

Units Affected
N/A
Distribution
Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.
Location
Birmingham, N/A

Frequently Asked Questions

What product was recalled?
Human lgG4 Subclass Liquid Reagent Kits For use on the Roche Cobas"6000 This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents Product Code LK009.CB. Recalled by The Binding Site Group, Ltd.. Units affected: N/A.
Why was this product recalled?
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 1, 2017. Severity: Moderate. Recall number: Z-1069-2017.