Severity
Critical
FDA Devices recall · Reported May 25, 2022
Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing p…
Atrium Medical Corporation recalled iCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 8540… - a critical-severity action.
iCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 8540… was recalled by Atrium Medical Corporation in May 25, 2022. Reason: Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub …. Check the official notice for the remedy. Verify recall #Z-1077-2022 with the FDA Devices before acting.
The recall
Atrium Medical Corporation issued this critical-severity FDA Devices recall-Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1077-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1077-2022) was formally reported on May 25, 2022, with the manufacturer initiating the action on March 3, 2022. It is classified under Critical severity (Class I), with a current status of Ongoing. Atrium Medical Corporation is listed as the recalling firm, operating out of Merrimack, NH. Federal records list the affected scope as 68812 units.
The documented reason for this recall is: Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administrat… Distribution data in the federal record shows the product reached: Foreign: Australia Austria Bahrain Belgium Bosnia-Herzegovina Brazil Bulgaria Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Finland France Germany …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
68812 units
Related Recalls
6
3 from same agency
iCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 85417 85418 85419 85420 85424 85440 85441 85442 85443 85444 85445 85450 85451 85452 85453 85454 85455 Product Code/UDI /Description: 85400 00650862854008 iCAST COVERED STENT,5MMX38MMX80CM 85401 00650862854015 iCAST COVERED STENT,5MMX59MMX80CM 85402 00650862854022 iCAST COVERED STENT,6MMX38MMX80CM 85403 00650862854039 iCAST COVERED STENT,6MMX59MMX80CM 85404 00650862854046 iCAST COVERED STENT,7MMX38MMX80CM 85405 00650862854053 iCAST COVERED STENT,7MMX59MMX80CM 85406 00650862854060 iCAST COVERED STENT,8MMX38MMX80CM 85407 00650862854077 iCAST COVERED STENT,8MMX59MMX80CM 85408 00650862854084 iCAST COVERED STENT,9MMX38MMX80CM 85409 00650862854091 iCAST COVERED STENT,9MMX59MMX80CM 85410 00650862854107 iCAST COVERED STENT,5MMX38MMX120CM 85411 00650862854114 iCAST COVERED STENT,5MMX59MMX120CM 85412 iCAST COVERED STENT,6MMX38MMX120CM 85413 00650862854138 iCAST COVERED STENT,6MMX59MMX120CM 85414 00650862854145 iCAST COVERED STENT,7MMX38MMX120CM 85415 00650862854152 iCAST COVERED STENT,7MMX59MMX120CM 85416 00650862854169 iCAST COVERED STENT,8MMX38MMX120CM 85417 00650862854176 iCAST COVERED STENT,8MMX59MMX120CM 85418 00650862854183 iCAST COVERED STENT,9MMX38MMX120CM 85419 00650862854190 iCAST COVERED STENT,9MMX59MMX120CM 85420 00650862854206 iCAST COVERED STENT,10MMX38MMX80CM 85424 00650862854244 iCAST COVERED STENT,10MMX38MMX120CM 85440 00650862854404 iCAST COVERED STENT,5MMX16MMX80CM 85441 00650862854411 iCAST COVERED STENT,5MMX22MMX80CM 85442 00650862854428 iCAST COVERED STENT 6MMX16MMX80CM 85443 00650862854435 iCAST COVERED STENT,6MMX22MMX80CM 85444 00650862854442 iCAST COVERED STENT,7MMX16MMX80CM 85445 00650862854459 iCAST COVERED STENT,7MMX22MMX80CM 85450 00650862854503 iCAST COVERED STENT,5MMX16MMX120CM 85451 00650862854510 iCAST COVERED STENT,5MMX22MMX120CM 85452 00650862854527 iCAST COVERED STENT,6MMX16MMX120CM 85453 00650862854534 iCAST COVERED STENT,6MMX22MMX120CM 85454 00650862854541 iCAST COVERED STENT,7MMX16MMX120CM 85455 00650862854558 iCAST COVERED STENT,7MMX22MMX120CM
Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administration of additional anesthesia and contrast which may impact patients' renal function. Potential harms include occlusion or embolism and associated response, with specific outcomes such as amputation, embolism, loss of organ function, organ infarction, or tissue infarction. ***Updated June 2023*** Additional surgical stress caused by prolonged interventional surgery has the potential to lead to myocardial infarction or death.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Ongoing |
| Recall number | Z-1077-2022 |
| Date reported | May 25, 2022 |
| Date initiated | March 3, 2022 |
| Recalling firm | Atrium Medical Corporation |
| Firm location | Merrimack, NH |
| Affected scope | 68812 units |
| Distribution | Foreign: Australia Austria Bahrain Belgium Bosnia-Herzegovina Brazil Bulgaria Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Finland France Germany Great Britain Gree… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 25, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.