Severity
Critical
Atrium Medical Corporation issued this FDA Devices recall on May 25, 2022. Classified as Critical severity (Class I). Approximately 68812 units units are affected. The recall was issued because: Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the d…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.
This FDA Devices action (record #Z-1077-2022) was formally reported on May 25, 2022, with the manufacturer initiating the action on March 3, 2022. It is classified under Critical severity (Class I), with a current status of Ongoing. Atrium Medical Corporation is listed as the recalling firm, operating out of Merrimack, NH. Federal records indicate 68812 units units are affected.
The documented reason for this recall is: Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administrat… Distribution data in the federal record shows the product reached: Foreign: Australia Austria Bahrain Belgium Bosnia-Herzegovina Brazil Bulgaria Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Finland France Germany …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 4 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.
Severity
Critical
Units Affected
68812 units
Related Recalls
6
6 from same agency
iCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 85417 85418 85419 85420 85424 85440 85441 85442 85443 85444 85445 85450 85451 85452 85453 85454 85455 Product Code/UDI /Description: 85400 00650862854008 iCAST COVERED STENT,5MMX38MMX80CM 85401 00650862854015 iCAST COVERED STENT,5MMX59MMX80CM 85402 00650862854022 iCAST COVERED STENT,6MMX38MMX80CM 85403 00650862854039 iCAST COVERED STENT,6MMX59MMX80CM 85404 00650862854046 iCAST COVERED STENT,7MMX38MMX80CM 85405 00650862854053 iCAST COVERED STENT,7MMX59MMX80CM 85406 00650862854060 iCAST COVERED STENT,8MMX38MMX80CM 85407 00650862854077 iCAST COVERED STENT,8MMX59MMX80CM 85408 00650862854084 iCAST COVERED STENT,9MMX38MMX80CM 85409 00650862854091 iCAST COVERED STENT,9MMX59MMX80CM 85410 00650862854107 iCAST COVERED STENT,5MMX38MMX120CM 85411 00650862854114 iCAST COVERED STENT,5MMX59MMX120CM 85412 iCAST COVERED STENT,6MMX38MMX120CM 85413 00650862854138 iCAST COVERED STENT,6MMX59MMX120CM 85414 00650862854145 iCAST COVERED STENT,7MMX38MMX120CM 85415 00650862854152 iCAST COVERED STENT,7MMX59MMX120CM 85416 00650862854169 iCAST COVERED STENT,8MMX38MMX120CM 85417 00650862854176 iCAST COVERED STENT,8MMX59MMX120CM 85418 00650862854183 iCAST COVERED STENT,9MMX38MMX120CM 85419 00650862854190 iCAST COVERED STENT,9MMX59MMX120CM 85420 00650862854206 iCAST COVERED STENT,10MMX38MMX80CM 85424 00650862854244 iCAST COVERED STENT,10MMX38MMX120CM 85440 00650862854404 iCAST COVERED STENT,5MMX16MMX80CM 85441 00650862854411 iCAST COVERED STENT,5MMX22MMX80CM 85442 00650862854428 iCAST COVERED STENT 6MMX16MMX80CM 85443 00650862854435 iCAST COVERED STENT,6MMX22MMX80CM 85444 00650862854442 iCAST COVERED STENT,7MMX16MMX80CM 85445 00650862854459 iCAST COVERED STENT,7MMX22MMX80CM 85450 00650862854503 iCAST COVERED STENT,5MMX16MMX120CM 85451 00650862854510 iCAST COVERED STENT,5MMX22MMX120CM 85452 00650862854527 iCAST COVERED STENT,6MMX16MMX120CM 85453 00650862854534 iCAST COVERED STENT,6MMX22MMX120CM 85454 00650862854541 iCAST COVERED STENT,7MMX16MMX120CM 85455 00650862854558 iCAST COVERED STENT,7MMX22MMX120CM
Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administration of additional anesthesia and contrast which may impact patients' renal function. Potential harms include occlusion or embolism and associated response, with specific outcomes such as amputation, embolism, loss of organ function, organ infarction, or tissue infarction. ***Updated June 2023*** Additional surgical stress caused by prolonged interventional surgery has the potential to lead to myocardial infarction or death.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Ongoing |
| Recall number | Z-1077-2022 |
| Date reported | May 25, 2022 |
| Date initiated | March 3, 2022 |
| Recalling firm | Atrium Medical Corporation |
| Units affected | 68812 units |
| Distribution | Foreign: Australia Austria Bahrain Belgium Bosnia-Herzegovina Brazil Bulgaria Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Finland France Germany Great Britain Gree… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
Learn how the US recall system works and how to protect yourself and your household.
How the US Recall System Works
The three federal agencies, how recalls are initiated, and what happens next.
Understanding Recall Severity Classes
What Class I, II, and III mean and which recalls demand immediate action.
What to Do When a Product Is Recalled
Verify, claim your remedy, report injuries, and navigate the process.
How to Check If Your Products Are Recalled
Step-by-step guide to checking food, products, medications, and vehicles.
Recalled Products in Your Home
A room-by-room household audit guide for active recalls.
Most Recalled Product Categories
Rankings of highest-recall categories from FDA, CPSC, and NHTSA.
Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as critical severity, meaning the product carries a reasonable probability of serious adverse health consequences or death. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.
Explore related product safety and public health data from federal sources.
Food Safety Inspections
FDA and state food facility inspection results — violations, enforcement actions, and compliance history for food manufacturers and processors.
PlainFoodSafe →
Ingredient Safety Data
FDA food additive and ingredient safety database — regulatory status, usage limits, and safety assessments for thousands of ingredients.
PlainIngredients →
Nutrition Data
USDA nutrition facts for 2M+ food items — calories, macros, vitamins, and ingredient analysis to verify what you consume.
GetFoodFacts →
Drug Safety Information
FDA drug data for 680+ medications — interactions, alternatives, side effects, and safety information for recalled and active drugs.
PlainMeds →
Product Injury Data
CPSC emergency room injury data for 838 product categories — 7.3M NEISS records tracking real-world consumer product injuries.
PlainSafety →
Recall Checker
Search our database of 83K+ recalls by product name, brand, or recall number across all agencies.
Check Now →
Recall Radar
Live feed of the latest recalls across FDA, CPSC, NHTSA, and USDA — filter by agency and severity.
View Radar →
Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.
Waldemar Link GmbH & Co. KG (Mfg Site) · 2026-03-18
Penner Patient Care, Inc. · 2026-03-18
Instrumentation Laboratory · 2026-03-18
Beckman Coulter Mishima K.K. · 2026-03-18
Trividia Health, Inc. · 2026-03-18
Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →
Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Read our methodology — how this data is sourced, computed, and verified.
Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).