FDA Devices recall · Reported April 10, 2019

UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. System Name/ Catalog Number: DxC 660i/ A64871, DxC 680i /A64903, DxC 860i /A64935, DxC 880i /A59102

When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandl…

Recall #
Z-1078-2019
Affected scope
0
Initiated
December 5, 2018
Verify with FDA Devices →
View my saved recalls

Beckman Coulter Inc. recalled UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-confi… - a moderate-severity action.

UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-confi… was recalled by Beckman Coulter Inc. in April 10, 2019. Reason: When racks are front-loaded onto the sample presentation unit while samples from the automation line are bein…. Check the official notice for the remedy. Verify recall #Z-1078-2019 with the FDA Devices before acting.

The recall

Beckman Coulter Inc. issued this moderate-severity FDA Devices recall-When racks are front-loaded onto the sample presentation unit while samples from the automation line are bein….

Moderate
severity level
Class II
classification
April 10, 2019
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1078-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. System Name/ Catalog Number: DxC 660i/ A64871, DxC 680i /A64903, DxC 860i /A64935, DxC 880i /A59102

Reason for recall

When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1078-2019
Date reported April 10, 2019
Date initiated December 5, 2018
Recalling firm Beckman Coulter Inc.
Firm location Brea, CA
Affected scope 0
Distribution US: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, Italy, Malaysia, Moro…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1078-2019) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. System Name/ Catalog Number: DxC 660i/ A64871, DxC 680i /A64903, DxC 860i /A64935, DxC 880i /A59102. Recalled by Beckman Coulter Inc.. Units affected: 0.
Why was this product recalled?
When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 10, 2019. Severity: Moderate. Recall number: Z-1078-2019.
Where was the recalled product distributed?
Distribution: US: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, Italy, Malaysia, Morocco, Netherlands, Nigeria, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Thailand, Tunisia, United Kingdom, United States, Vietnam.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported April 10, 2019. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026