Severity
Moderate
FDA Devices recall · Reported April 10, 2019
When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandl…
Beckman Coulter Inc. recalled UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-confi… - a moderate-severity action.
UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-confi… was recalled by Beckman Coulter Inc. in April 10, 2019. Reason: When racks are front-loaded onto the sample presentation unit while samples from the automation line are bein…. Check the official notice for the remedy. Verify recall #Z-1078-2019 with the FDA Devices before acting.
The recall
Beckman Coulter Inc. issued this moderate-severity FDA Devices recall-When racks are front-loaded onto the sample presentation unit while samples from the automation line are bein….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1078-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1078-2019) was formally reported on April 10, 2019, with the manufacturer initiating the action on December 5, 2018. It is classified under Moderate severity (Class II), with a current status of Ongoing. Beckman Coulter Inc. is listed as the recalling firm, operating out of Brea, CA. Federal records list the affected scope as 0.
The documented reason for this recall is: When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an… Distribution data in the federal record shows the product reached: US: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, I…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
0
Related Recalls
6
3 from same agency
UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. System Name/ Catalog Number: DxC 660i/ A64871, DxC 680i /A64903, DxC 860i /A64935, DxC 880i /A59102
When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1078-2019 |
| Date reported | April 10, 2019 |
| Date initiated | December 5, 2018 |
| Recalling firm | Beckman Coulter Inc. |
| Firm location | Brea, CA |
| Affected scope | 0 |
| Distribution | US: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, Italy, Malaysia, Moro… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 10, 2019. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.