MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Reported: January 21, 2026 Initiated: December 10, 2025 #Z-1078-2026

Product Description

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Reason for Recall

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Details

Recalling Firm
Elekta, Inc.
Units Affected
1
Distribution
US Nationwide distribution in the state of TX.
Location
Atlanta, GA

Frequently Asked Questions

What product was recalled?
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO). Recalled by Elekta, Inc.. Units affected: 1.
Why was this product recalled?
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice