PlainRecalls
FDA Devices Moderate Class II Terminated

RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Reported: March 12, 2014 Initiated: November 25, 2013 #Z-1081-2014

Product Description

RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Reason for Recall

During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.

Details

Recalling Firm
Elekta, Inc.
Units Affected
489
Distribution
Worldwide Distribution - USA including AL, CA, CT, GA, FL, IL, IN, KS, KY, LA, MD, MA, MI, MN, MO, NJ, NY, OH, PA, TN, TX, VA, and Internationally to Argentina, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cuba, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherland, Nigeria, Pakistan, Panama, Peru, Philippines, Portugal, Russia, Serbia, Solvenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Vietnam, and Yemen.
Location
Atlanta, GA

Frequently Asked Questions

What product was recalled?
RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.. Recalled by Elekta, Inc.. Units affected: 489.
Why was this product recalled?
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 12, 2014. Severity: Moderate. Recall number: Z-1081-2014.