Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene

Reported: February 17, 2021 Initiated: December 22, 2020 #Z-1099-2021

Product Description

Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene

Reason for Recall

Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.

Details

Recalling Firm
Cytocell Ltd.
Units Affected
72
Distribution
Domestic distribution to California, Illinois, Michigan, Minnesota, New Jersey, Texas, and Washington. Foreign distribution to Canada.
Location
Cambridge, N/A

Frequently Asked Questions

What product was recalled?
Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene. Recalled by Cytocell Ltd.. Units affected: 72.
Why was this product recalled?
Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 17, 2021. Severity: Moderate. Recall number: Z-1099-2021.