Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system syngo(R) Workflow SLR is a digital radiology information system (RIS) with integrated modules for patient administration, examination, reporting, statistics and system administration, as well as multiple interfaces for patient registration, order entry, and external billing modules. It is not intended to treat disease or administer any medicinal substances or drugs. There is no conta

Reported: April 24, 2013 Initiated: October 5, 2012 #Z-1105-2013

Product Description

Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system syngo(R) Workflow SLR is a digital radiology information system (RIS) with integrated modules for patient administration, examination, reporting, statistics and system administration, as well as multiple interfaces for patient registration, order entry, and external billing modules. It is not intended to treat disease or administer any medicinal substances or drugs. There is no contact between device and patient. syngo(R) Workflow SLR can be used as a separate RIS, interfacing to a Hospital Information System (HIS) or Picture Archiving and Communication Systems (PACS

Reason for Recall

Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when the Interactive Document is viewed or auto displayed, dismissed by the user, and then a different patient and signed exam is selected in single patient view. The Notes Viewer will

Details

Units Affected
37
Distribution
Worldwide Distribution - USA including the states of AL, CA, CO, DC, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WI, WV, and WY. and the countries of Canada and Australia.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system syngo(R) Workflow SLR is a digital radiology information system (RIS) with integrated modules for patient administration, examination, reporting, statistics and system administration, as well as multiple interfaces for patient registration, order entry, and external billing modules. It is not intended to treat disease or administer any medicinal substances or drugs. There is no contact between device and patient. syngo(R) Workflow SLR can be used as a separate RIS, interfacing to a Hospital Information System (HIS) or Picture Archiving and Communication Systems (PACS. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 37.
Why was this product recalled?
Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when the Interactive Document is viewed or auto displayed, dismissed by the user, and then a different patient and signed exam is selected in single patient view. The Notes Viewer will
Which agency issued this recall?
This recall was issued by the FDA Devices on April 24, 2013. Severity: Moderate. Recall number: Z-1105-2013.