ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar).

Reported: April 24, 2013 Initiated: March 6, 2013 #Z-1113-2013

Product Description

ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar).

Reason for Recall

Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.

Details

Units Affected
Both devices 4467
Distribution
Nationwide Distribution.