VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198
Reported: May 25, 2022 Initiated: March 25, 2022 #Z-1113-2022
Product Description
VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198
Reason for Recall
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.
Details
- Recalling Firm
- Ortho-Clinical Diagnostics, Inc
- Units Affected
- 1575 units total: 1183 US; 392 OUS
- Distribution
- Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia Singapore Spain Sweden United Kingdom
- Location
- Raritan, NJ
Frequently Asked Questions
What product was recalled? ▼
VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198. Recalled by Ortho-Clinical Diagnostics, Inc. Units affected: 1575 units total: 1183 US; 392 OUS.
Why was this product recalled? ▼
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 25, 2022. Severity: Moderate. Recall number: Z-1113-2022.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11