Severity
Moderate
FDA Devices recall · Reported April 24, 2013
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient id…
Roche Diagnostics Operations, Inc. recalled CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the Co… - a moderate-severity action.
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the Co… was recalled by Roche Diagnostics Operations, Inc. in April 24, 2013. Reason: Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading met…. Check the official notice for the remedy. Verify recall #Z-1114-2013 with the FDA Devices before acting.
The recall
Roche Diagnostics Operations, Inc. issued this moderate-severity FDA Devices recall-Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading met….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1114-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1114-2013) was formally reported on April 24, 2013, with the manufacturer initiating the action on March 6, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Operations, Inc. is listed as the recalling firm, operating out of Indianapolis, IN. Federal records list the affected scope as 4467 both devices.
The documented reason for this recall is: Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail… Distribution data in the federal record shows the product reached: Nationwide Distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4467 both devices
Related Recalls
6
3 from same agency
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1114-2013 |
| Date reported | April 24, 2013 |
| Date initiated | March 6, 2013 |
| Recalling firm | Roche Diagnostics Operations, Inc. |
| Firm location | Indianapolis, IN |
| Affected scope | 4467 both devices |
| Distribution | Nationwide Distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 24, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.