IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 10706293 (US Kit Catalog) Product Usage: For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the man
Negative bias results with the IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit was confirmed for two lots (D0119 and D0120). The investigation by S…
Siemens Healthcare Diagnostics, Inc. recalled IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 1070… - a moderate-severity action.
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 1070… was recalled by Siemens Healthcare Diagnostics, Inc. in February 8, 2017. Reason: Negative bias results with the IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit…. Check the official notice for the remedy. Verify recall #Z-1120-2017 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Negative bias results with the IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit….
- Moderate
- severity level
- Class II
- classification
- February 8, 2017
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1120-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 10706293 (US Kit Catalog) Product Usage: For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.
Reason for recall
Negative bias results with the IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit was confirmed for two lots (D0119 and D0120). The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1120-2017 |
| Date reported | February 8, 2017 |
| Date initiated | December 21, 2016 |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Firm location | Tarrytown, NY |
| Affected scope | 1379 kits (190 kits in U.S.A and 1189 Kits internationally) in total |
| Distribution | Worldwide Distribution - US Nationwide and the countries of Denmark, Brazil, Croatia and Germany |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1120-2017) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026