PlainRecalls
ModerateClass IIOngoing

FDA Devices recall · Reported January 28, 2026

ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is deliver…

Recall #
Z-1120-2026
Initiated
December 15, 2025
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Icu Medical, Inc. recalled ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philip… - a moderate-severity action.

ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philip… was recalled by Icu Medical, Inc. in January 28, 2026. Reason: Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream …. Check the official notice for the remedy. Verify recall #Z-1120-2026 with the FDA Devices before acting.

The recall

Icu Medical, Inc. issued this moderate-severity FDA Devices recall-Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream ….

Moderate
severity level
Class II
classification
January 28, 2026
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1120-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1120-2026) was formally reported on January 28, 2026, with the manufacturer initiating the action on December 15, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Icu Medical, Inc. is listed as the recalling firm, operating out of Lake Forest, IL. The number of affected units is not disclosed in the agency filing, which is common for drug and food recalls where lot-level tracking supersedes unit counts.

The documented reason for this recall is: Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at th… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the country of Philippines.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

Unknown

Related Recalls

6

3 from same agency

Product description

ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Reason for recall

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1120-2026
Date reported January 28, 2026
Date initiated December 15, 2025
Recalling firm Icu Medical, Inc.
Firm location Lake Forest, IL
Affected scope Not disclosed
Distribution Worldwide distribution - US Nationwide and the country of Philippines.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1120-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines). Recalled by Icu Medical, Inc..
Why was this product recalled?
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 28, 2026. Severity: Moderate. Recall number: Z-1120-2026.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide and the country of Philippines..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1120-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 28, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.