FDA Devices Moderate Class II Terminated

Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.

Reported: February 25, 2015 Initiated: January 26, 2015 #Z-1130-2015

Product Description

Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.

Reason for Recall

Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter.

Details

Recalling Firm
Haemonetics Corporation
Units Affected
4,143 kits
Distribution
Worldwide Distribution -- US, Canada, Czech Republic, Denmark, Great Britain, and Sweden.
Location
Braintree, MA

Frequently Asked Questions

What product was recalled?
Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.. Recalled by Haemonetics Corporation. Units affected: 4,143 kits.
Why was this product recalled?
Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 25, 2015. Severity: Moderate. Recall number: Z-1130-2015.