Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected PCs: FlexViewing PC Azurion Model Name - Model Number: Azurion 7 B12 722067; Azurion 7 B20 722068; Azurion 7 M12 722078 722223; Azurion 7 M20 722079;

Reported: February 28, 2024 Initiated: January 2, 2024 #Z-1145-2024

Product Description

Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected PCs: FlexViewing PC Azurion Model Name - Model Number: Azurion 7 B12 722067; Azurion 7 B20 722068; Azurion 7 M12 722078 722223; Azurion 7 M20 722079;

Reason for Recall

Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.

Details

Units Affected
1,993 US; 2,662 ROW
Distribution
Worldwide - US Nationwide distribution.
Location
Best, N/A

Frequently Asked Questions

What product was recalled?
Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected PCs: FlexViewing PC Azurion Model Name - Model Number: Azurion 7 B12 722067; Azurion 7 B20 722068; Azurion 7 M12 722078 722223; Azurion 7 M20 722079;. Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. Units affected: 1,993 US; 2,662 ROW.
Why was this product recalled?
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 28, 2024. Severity: Moderate. Recall number: Z-1145-2024.