J-Plasma(R) Handpiece, Catalog Numbers: GS-018C, GS-270C, BVX-270B

Reported: March 28, 2018 Initiated: February 8, 2018 #Z-1155-2018

Product Description

J-Plasma(R) Handpiece, Catalog Numbers: GS-018C, GS-270C, BVX-270B

Reason for Recall

Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.

Details

Recalling Firm
Bovie Medical Corporation
Units Affected
5416 units
Distribution
worldwide
Location
Clearwater, FL

Frequently Asked Questions

What product was recalled?
J-Plasma(R) Handpiece, Catalog Numbers: GS-018C, GS-270C, BVX-270B. Recalled by Bovie Medical Corporation. Units affected: 5416 units.
Why was this product recalled?
Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.