Severity
Moderate
FDA Devices recall · Reported March 19, 2014
Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft …
Smith & Nephew Inc. recalled Smith & Nephew RENASYS-Gauze with Soft Port Kit REF #'s: 66800933, 66800934, 66800935, … - a moderate-severity action.
Smith & Nephew RENASYS-Gauze with Soft Port Kit REF #'s: 66800933, 66800934, 66800935, … was recalled by Smith & Nephew Inc. in March 19, 2014. Reason: Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blo…. Check the official notice for the remedy. Verify recall #Z-1160-2014 with the FDA Devices before acting.
The recall
Smith & Nephew Inc. issued this moderate-severity FDA Devices recall-Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blo….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1160-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1160-2014) was formally reported on March 19, 2014, with the manufacturer initiating the action on January 9, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Smith & Nephew Inc. is listed as the recalling firm, operating out of Saint Petersburg, FL. Federal records list the affected scope as 498,980 individuals units.
The documented reason for this recall is: Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improp… Distribution data in the federal record shows the product reached: Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries D…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
498,980 individuals units
Related Recalls
6
3 from same agency
Smith & Nephew RENASYS-Gauze with Soft Port Kit REF #'s: 66800933, 66800934, 66800935, and 66800936 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1160-2014 |
| Date reported | March 19, 2014 |
| Date initiated | January 9, 2014 |
| Recalling firm | Smith & Nephew Inc. |
| Firm location | Saint Petersburg, FL |
| Affected scope | 498,980 individuals units |
| Distribution | Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argen… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 19, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.