Luminos Agile Max with software VE10 & VF11- A diagnostic imaging system for radiographic and fluoroscopic studies Model Number: 10762472

Reported: March 1, 2023 Initiated: December 21, 2022 #Z-1165-2023

Product Description

Luminos Agile Max with software VE10 & VF11- A diagnostic imaging system for radiographic and fluoroscopic studies Model Number: 10762472

Reason for Recall

Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values

Details

Units Affected
28 units. Expanded Recall: 506 units
Distribution
US Nationwide distribution.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
Luminos Agile Max with software VE10 & VF11- A diagnostic imaging system for radiographic and fluoroscopic studies Model Number: 10762472. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 28 units. Expanded Recall: 506 units.
Why was this product recalled?
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
Which agency issued this recall?
This recall was issued by the FDA Devices on March 1, 2023. Severity: Moderate. Recall number: Z-1165-2023.