Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.

Reported: February 28, 2024 Initiated: January 3, 2024 #Z-1172-2024

Product Description

Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.

Reason for Recall

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Details

Units Affected
8
Distribution
US domestic distribution nationwide. International distribution to Israel.
Location
Lexington, MA

Frequently Asked Questions

What product was recalled?
Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.. Recalled by FUJIFILM Healthcare Americas Corporation. Units affected: 8.
Why was this product recalled?
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 28, 2024. Severity: Moderate. Recall number: Z-1172-2024.