Severity
Moderate
FDA Devices recall · Reported April 4, 2018
During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the scre…
Biomet, Inc. recalled Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Prox… - a moderate-severity action.
Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Prox… was recalled by Biomet, Inc. in April 4, 2018. Reason: During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with…. Check the official notice for the remedy. Verify recall #Z-1173-2018 with the FDA Devices before acting.
The recall
Biomet, Inc. issued this moderate-severity FDA Devices recall-During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1173-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1173-2018) was formally reported on April 4, 2018, with the manufacturer initiating the action on February 28, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Biomet, Inc. is listed as the recalling firm, operating out of Warsaw, IN. Federal records list the affected scope as 634.
The documented reason for this recall is: During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Cond… Distribution data in the federal record shows the product reached: Distribution US Nationwide. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
634
Related Recalls
6
3 from same agency
Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral Reconstruction, and Total Humeral Reconstruction
During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1173-2018 |
| Date reported | April 4, 2018 |
| Date initiated | February 28, 2018 |
| Recalling firm | Biomet, Inc. |
| Firm location | Warsaw, IN |
| Affected scope | 634 |
| Distribution | Distribution US Nationwide |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 4, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.