PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported March 19, 2014

Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially infectious. IVD; The Cannabinoids (cTHC also known as THC version c) DAU negative and positive controls are human urine-based liquids and ready to use. The constituent is a processed drug-free human urine matrix, containing buffers, stabilizers, and less than 0.09 % of sodium azide. The controls are prepared by spiking known concentrations of 11-no

Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards…

Recall #
Z-1178-2014
Affected scope
130 items
Initiated
February 6, 2014
Verify with FDA Devices →
View my saved recalls

Lin-Zhi International Inc recalled Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared fro… - a moderate-severity action.

Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared fro… was recalled by Lin-Zhi International Inc in March 19, 2014. Reason: Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 o…. Check the official notice for the remedy. Verify recall #Z-1178-2014 with the FDA Devices before acting.

The recall

Lin-Zhi International Inc issued this moderate-severity FDA Devices recall-Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 o….

Moderate
severity level
130 units
affected scope
Class II
classification
March 19, 2014
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1178-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially infectious. IVD; The Cannabinoids (cTHC also known as THC version c) DAU negative and positive controls are human urine-based liquids and ready to use. The constituent is a processed drug-free human urine matrix, containing buffers, stabilizers, and less than 0.09 % of sodium azide. The controls are prepared by spiking known concentrations of 11-nor-9-THC-9-COOH into the drug-free matrix. cTHC is referenced as an abbreviation for 11-nor-9-THC-9-COOH and it may be referred to simply as THC in the text of this report. The controls contain 37.5 ng/mL and 62.5 ng/mL of cTHC,¿ 10% by GC/MS or LC/MS analysis. Lin Zhi International inc. Sunnyvale, CA 94085 The Cannabinoid Drug of Abuse (DAU) negative and positive controls are used as assay quality control material to monitor the precision of the Lin-Zhi International, Inc. (LZI) Cannabinoid Enzyme Immunoassay (Ref # 0070c) on a number of automated clinical chemistry analyzers.

Reason for recall

Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards the end of their shelf life due to use of dated but unexpired secondary stock used to create the products.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1178-2014
Date reported March 19, 2014
Date initiated February 6, 2014
Recalling firm Lin-Zhi International Inc
Firm location Sunnyvale, CA
Affected scope 130 items
Distribution US Distribution including the states of SC, NC, LA, FL, TX,, AL and GA and Internationally to Germany.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

130 items units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1178-2014) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially infectious. IVD; The Cannabinoids (cTHC also known as THC version c) DAU negative and positive controls are human urine-based liquids and ready to use. The constituent is a processed drug-free human urine matrix, containing buffers, stabilizers, and less than 0.09 % of sodium azide. The controls are prepared by spiking known concentrations of 11-nor-9-THC-9-COOH into the drug-free matrix. cTHC is referenced as an abbreviation for 11-nor-9-THC-9-COOH and it may be referred to simply as THC in the text of this report. The controls contain 37.5 ng/mL and 62.5 ng/mL of cTHC,¿ 10% by GC/MS or LC/MS analysis. Lin Zhi International inc. Sunnyvale, CA 94085 The Cannabinoid Drug of Abuse (DAU) negative and positive controls are used as assay quality control material to monitor the precision of the Lin-Zhi International, Inc. (LZI) Cannabinoid Enzyme Immunoassay (Ref # 0070c) on a number of automated clinical chemistry analyzers.. Recalled by Lin-Zhi International Inc. Units affected: 130 items.
Why was this product recalled?
Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards the end of their shelf life due to use of dated but unexpired secondary stock used to create the products.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 19, 2014. Severity: Moderate. Recall number: Z-1178-2014.
Where was the recalled product distributed?
Distribution: US Distribution including the states of SC, NC, LA, FL, TX,, AL and GA and Internationally to Germany..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported March 19, 2014. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026