PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported February 22, 2017

LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.

Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, w…

Recall #
Z-1191-2017
Affected scope
13 devices
Initiated
January 10, 2017
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Encore Medical, Lp recalled LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revisi… - a moderate-severity action.

LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revisi… was recalled by Encore Medical, Lp in February 22, 2017. Reason: Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal…. Check the official notice for the remedy. Verify recall #Z-1191-2017 with the FDA Devices before acting.

The recall

Encore Medical, Lp issued this moderate-severity FDA Devices recall-Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal….

Moderate
severity level
13 units
affected scope
Class II
classification
February 22, 2017
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1191-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1191-2017) was formally reported on February 22, 2017, with the manufacturer initiating the action on January 10, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Encore Medical, Lp is listed as the recalling firm, operating out of Austin, TX. Federal records list the affected scope as 13 devices.

The documented reason for this recall is: Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma … Distribution data in the federal record shows the product reached: US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

13 devices

Related Recalls

6

3 from same agency

Product description

LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.

Reason for recall

Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1191-2017
Date reported February 22, 2017
Date initiated January 10, 2017
Recalling firm Encore Medical, Lp
Firm location Austin, TX
Affected scope 13 devices
Distribution US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

13 devices units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1191-2017) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.. Recalled by Encore Medical, Lp. Units affected: 13 devices.
Why was this product recalled?
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 22, 2017. Severity: Moderate. Recall number: Z-1191-2017.
Where was the recalled product distributed?
Distribution: US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1191-2017) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 22, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.