Severity
Moderate
FDA Devices recall · Reported February 22, 2017
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, w…
Encore Medical, Lp recalled LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revisi… - a moderate-severity action.
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revisi… was recalled by Encore Medical, Lp in February 22, 2017. Reason: Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal…. Check the official notice for the remedy. Verify recall #Z-1191-2017 with the FDA Devices before acting.
The recall
Encore Medical, Lp issued this moderate-severity FDA Devices recall-Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1191-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1191-2017) was formally reported on February 22, 2017, with the manufacturer initiating the action on January 10, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Encore Medical, Lp is listed as the recalling firm, operating out of Austin, TX. Federal records list the affected scope as 13 devices.
The documented reason for this recall is: Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma … Distribution data in the federal record shows the product reached: US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
13 devices
Related Recalls
6
3 from same agency
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1191-2017 |
| Date reported | February 22, 2017 |
| Date initiated | January 10, 2017 |
| Recalling firm | Encore Medical, Lp |
| Firm location | Austin, TX |
| Affected scope | 13 devices |
| Distribution | US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 22, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.