PlainRecalls
FDA Devices Verify with FDA Devices → Moderate Class II Terminated

LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.

Reported: February 22, 2017 Initiated: January 10, 2017 #Z-1191-2017 13 devices units

Encore Medical, Lp issued this FDA Devices recall on February 22, 2017. Classified as Moderate severity (Class II). Approximately 13 devices units are affected. The recall was issued because: Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-1191-2017) was formally reported on February 22, 2017, with the manufacturer initiating the action on January 10, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Encore Medical, Lp is listed as the recalling firm, operating out of Austin, TX. Federal records indicate 13 devices units are affected.

The documented reason for this recall is: Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma … Distribution data in the federal record shows the product reached: US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 9 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

13 devices

Related Recalls

6

6 from same agency

Product Description

LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.

Reason for Recall

Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.

Details

Recalling Firm
Encore Medical, Lp
Units Affected
13 devices
Distribution
US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI
Location
Austin, TX

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1191-2017
Date reported February 22, 2017
Date initiated January 10, 2017
Recalling firm Encore Medical, Lp
Units affected 13 devices
Distribution US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

13 devices units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.. Recalled by Encore Medical, Lp. Units affected: 13 devices.
Why was this product recalled?
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 22, 2017. Severity: Moderate. Recall number: Z-1191-2017.
Where was the recalled product distributed?
Distribution: US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1191-2017) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).