FDA Devices recall · Reported March 19, 2014
Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes inc
Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on …
Ge Healthcare, LLC recalled Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usa… - a moderate-severity action.
Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usa… was recalled by Ge Healthcare, LLC in March 19, 2014. Reason: Potential safety issue related to unintended radial detector motion, which may occur during patient setup or …. Check the official notice for the remedy. Verify recall #Z-1200-2014 with the FDA Devices before acting.
The recall
Ge Healthcare, LLC issued this moderate-severity FDA Devices recall-Potential safety issue related to unintended radial detector motion, which may occur during patient setup or ….
- Moderate
- severity level
- 3K units
- affected scope
- Class II
- classification
- March 19, 2014
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1200-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Elscint APEX Helix, dual-head, multi-purpose, Slip-Ring, digital gamma camera Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes include planar (Static, Multi-gated, Dynamic, Whole body scanning) and tomographic (SPECT, Gated SPECT, Whole body SPECT, Camera based PET). Acquisition types include single and multi-isotope/multi-peak frame/list mode single-photon and positron imaging. Optional imaging-enhancement features include assortment of collimators, gating by physiological signals, real-time automatic body contouring, and CT-based attenuation correction and functional anatomic mapping.
Reason for recall
Potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. This failure mode has been observed on a single system, in which the detector moved to the home position and then slowly starting drifting inward toward the patient. This generated both a visual and audio alert condition on both the console and the gantry. N
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1200-2014 |
| Date reported | March 19, 2014 |
| Date initiated | February 7, 2014 |
| Recalling firm | Ge Healthcare, LLC |
| Firm location | Waukesha, WI |
| Affected scope | 3062 units (1182 US, 1880 OUS) |
| Distribution | Worldwide Distribution-USA (nationwide) including Puerto Rico and Guam and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Domi… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1200-2014) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.