Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HI ONCOR Impression ONCOR Impression Plus ONCOR Avant Garde ONCOR Expression ARTISTE, ONCOR, and PRIMUS running the following: Control Console from software version 13.0.302 and higher; Control Console from software version 9.2.400 and higher; and Control Console from software version 11.0.400 and higher.

Reported: May 1, 2019 Initiated: March 20, 2019 #Z-1204-2019

Product Description

Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HI ONCOR Impression ONCOR Impression Plus ONCOR Avant Garde ONCOR Expression ARTISTE, ONCOR, and PRIMUS running the following: Control Console from software version 13.0.302 and higher; Control Console from software version 9.2.400 and higher; and Control Console from software version 11.0.400 and higher.

Reason for Recall

Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field Sequence (AFS) Automatic Motion Protection (AMP) function implemented at the control of the Digital Linear Accelerator.

Details

Units Affected
120 US units
Distribution
Nationwide
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HI ONCOR Impression ONCOR Impression Plus ONCOR Avant Garde ONCOR Expression ARTISTE, ONCOR, and PRIMUS running the following: Control Console from software version 13.0.302 and higher; Control Console from software version 9.2.400 and higher; and Control Console from software version 11.0.400 and higher.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 120 US units.
Why was this product recalled?
Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field Sequence (AFS) Automatic Motion Protection (AMP) function implemented at the control of the Digital Linear Accelerator.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 1, 2019. Severity: Moderate. Recall number: Z-1204-2019.