DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate

Reported: March 8, 2023 Initiated: January 30, 2023 #Z-1206-2023

Product Description

DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate Current (AC) to the Overhead Tube Crane (OTC).

Reason for Recall

After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury

Details

Recalling Firm
Carestream Health, Inc.
Units Affected
72 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.
Location
Rochester, NY

Frequently Asked Questions

What product was recalled?
DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate Current (AC) to the Overhead Tube Crane (OTC).. Recalled by Carestream Health, Inc.. Units affected: 72 units.
Why was this product recalled?
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury
Which agency issued this recall?
This recall was issued by the FDA Devices on March 8, 2023. Severity: Moderate. Recall number: Z-1206-2023.