Severity
Low
FDA Devices recall · Reported February 22, 2017
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that p…
Siemens Healthcare Diagnostics, Inc. recalled The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine … - a low-severity action.
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine … was recalled by Siemens Healthcare Diagnostics, Inc. in February 22, 2017. Reason: Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension V…. Check the official notice for the remedy. Verify recall #Z-1216-2017 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics, Inc. issued this low-severity FDA Devices recall-Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension V….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1216-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1216-2017) was formally reported on February 22, 2017, with the manufacturer initiating the action on May 5, 2016. It is classified under Low severity (Class III), with a current status of Terminated. Siemens Healthcare Diagnostics, Inc. is listed as the recalling firm, operating out of Newark, DE. Federal records list the affected scope as 13,147 units.
The documented reason for this recall is: Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstra… Distribution data in the federal record shows the product reached: Distributed throughout the United States. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified low severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Low
Affected scope
13,147 units
Related Recalls
6
3 from same agency
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Low (Class III) |
| Status | Terminated |
| Recall number | Z-1216-2017 |
| Date reported | February 22, 2017 |
| Date initiated | May 5, 2016 |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Firm location | Newark, DE |
| Affected scope | 13,147 units |
| Distribution | Distributed throughout the United States |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 22, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.