FDA Devices recall · Reported April 4, 2018
The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a pathway into the pedicle. The instrument has three types of buttons to remotely control the software: the UP, DOWN and SELECT buttons. The Pedicle Feeler is laser marked in 10 mm intervals to help indicate the depth in which the feeler tip has been inserted into the pedicle. Next to the navigational information, these depth indicators on the feeler tip are a
Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ N…
Stryker Instruments Div. of Stryker Corporation recalled The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control… - a moderate-severity action.
The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control… was recalled by Stryker Instruments Div. of Stryker Corporation in April 4, 2018. Reason: Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Tho…. Check the official notice for the remedy. Verify recall #Z-1216-2018 with the FDA Devices before acting.
The recall
Stryker Instruments Div. of Stryker Corporation issued this moderate-severity FDA Devices recall-Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Tho….
- Moderate
- severity level
- 28 units
- affected scope
- Class II
- classification
- April 4, 2018
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1216-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a pathway into the pedicle. The instrument has three types of buttons to remotely control the software: the UP, DOWN and SELECT buttons. The Pedicle Feeler is laser marked in 10 mm intervals to help indicate the depth in which the feeler tip has been inserted into the pedicle. Next to the navigational information, these depth indicators on the feeler tip are also helpful in determining the appropriate pedicle screw length.
Reason for recall
Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ Navigation System because the settings on the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1216-2018 |
| Date reported | April 4, 2018 |
| Date initiated | October 16, 2017 |
| Recalling firm | Stryker Instruments Div. of Stryker Corporation |
| Firm location | Portage, MI |
| Affected scope | 28 |
| Distribution | International , US and US Govt. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1216-2018) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026