PlainRecalls
FDA Devices Moderate Class II Ongoing

A.L.P.S. mvX- MEDIAL TIBIA PLATE 6H RT-Medial Tibia Plate, 6-Hole, Right Model/Catalog Number: 770715062 Intended for Fixation of fractures of the distal tibia

Reported: March 5, 2025 Initiated: January 21, 2025 #Z-1220-2025

Product Description

A.L.P.S. mvX- MEDIAL TIBIA PLATE 6H RT-Medial Tibia Plate, 6-Hole, Right Model/Catalog Number: 770715062 Intended for Fixation of fractures of the distal tibia

Reason for Recall

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Details

Recalling Firm
Tyber Medical
Units Affected
210 units
Distribution
Nationwide
Location
Bethlehem, PA

Frequently Asked Questions

What product was recalled?
A.L.P.S. mvX- MEDIAL TIBIA PLATE 6H RT-Medial Tibia Plate, 6-Hole, Right Model/Catalog Number: 770715062 Intended for Fixation of fractures of the distal tibia. Recalled by Tyber Medical. Units affected: 210 units.
Why was this product recalled?
Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Which agency issued this recall?
This recall was issued by the FDA Devices on March 5, 2025. Severity: Moderate. Recall number: Z-1220-2025.