A.L.P.S. mvX- MEDIAL TIBIA PLATE 10H LT- Medial Tibia Plate, 10-Hole, Left. Intended Fixation of fractures of the distal tibia Model/Catalog Number: 770715101
Reported: March 5, 2025 Initiated: January 21, 2025 #Z-1221-2025
Product Description
A.L.P.S. mvX- MEDIAL TIBIA PLATE 10H LT- Medial Tibia Plate, 10-Hole, Left. Intended Fixation of fractures of the distal tibia Model/Catalog Number: 770715101
Reason for Recall
Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Details
- Recalling Firm
- Tyber Medical
- Units Affected
- 200 units
- Distribution
- Nationwide
- Location
- Bethlehem, PA
Frequently Asked Questions
What product was recalled? ▼
A.L.P.S. mvX- MEDIAL TIBIA PLATE 10H LT- Medial Tibia Plate, 10-Hole, Left. Intended Fixation of fractures of the distal tibia Model/Catalog Number: 770715101. Recalled by Tyber Medical. Units affected: 200 units.
Why was this product recalled? ▼
Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on March 5, 2025. Severity: Moderate. Recall number: Z-1221-2025.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11