VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents

Reported: May 1, 2019 Initiated: March 11, 2019 #Z-1222-2019

Product Description

VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents

Reason for Recall

One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.

Details

Units Affected
80
Distribution
Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, IL, OH, TX, and VA. The products were distributed to the following foreign countries: Australia, Belgium, Canada, Chile, Columbia, France, India, Italy, Japan, Portugal, Singapore, Spain, United Kingdom
Location
Rochester, NY

Frequently Asked Questions

What product was recalled?
VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents. Recalled by Ortho-Clinical Diagnostics. Units affected: 80.
Why was this product recalled?
One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 1, 2019. Severity: Moderate. Recall number: Z-1222-2019.