Severity
Moderate
FDA Devices recall · Reported February 11, 2026
Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Un…
Intuitive Surgical, Inc. recalled Brand Name: Da Vinci Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgica… - a moderate-severity action.
Brand Name: Da Vinci Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgica… was recalled by Intuitive Surgical, Inc. in February 11, 2026. Reason: Due to a software implementation error that allowed instruments arms failing an important diagnostic test to …. Check the official notice for the remedy. Verify recall #Z-1232-2026 with the FDA Devices before acting.
The recall
Intuitive Surgical, Inc. issued this moderate-severity FDA Devices recall-Due to a software implementation error that allowed instruments arms failing an important diagnostic test to ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1232-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1232-2026) was formally reported on February 11, 2026, with the manufacturer initiating the action on November 7, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Intuitive Surgical, Inc. is listed as the recalling firm, operating out of Sunnyvale, CA. Federal records list the affected scope as 219.
The documented reason for this recall is: Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion a… Distribution data in the federal record shows the product reached: U.S.: AL, AR, CA, CO, CT, FL, ID, IL, IN, KS, KY, MD, MI, MN, MS, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. O.U.S.: Austria, Brazil, China, France, Germany, Greece, India, Italy, Japan, Lebanon, N…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
219
Related Recalls
6
3 from same agency
Brand Name: Da Vinci Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgical System Model/Catalog Number: IS4000; IS4200; IS5000 Software Version: N/A Product Description: The da Vinci Xi, X, and da Vinci 5 surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1232-2026 |
| Date reported | February 11, 2026 |
| Date initiated | November 7, 2025 |
| Recalling firm | Intuitive Surgical, Inc. |
| Firm location | Sunnyvale, CA |
| Affected scope | 219 |
| Distribution | U.S.: AL, AR, CA, CO, CT, FL, ID, IL, IN, KS, KY, MD, MI, MN, MS, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. O.U.S.: Austria, Brazil, China, France, Germany, Greece, India, Italy, Japan, Lebanon, Netherlands, South Ko… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 11, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.