PlainRecalls
FDA Devices Moderate Class II Ongoing

VITROS XT 7600 Integrated System, Catalog No. 6844461

Reported: March 13, 2024 Initiated: January 26, 2024 #Z-1233-2024

Product Description

VITROS XT 7600 Integrated System, Catalog No. 6844461

Reason for Recall

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

Details

Units Affected
142 US; 78 OUS
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Location
Rochester, NY

Frequently Asked Questions

What product was recalled?
VITROS XT 7600 Integrated System, Catalog No. 6844461. Recalled by Ortho-Clinical Diagnostics, Inc.. Units affected: 142 US; 78 OUS.
Why was this product recalled?
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 13, 2024. Severity: Moderate. Recall number: Z-1233-2024.