FDA Devices recall · Reported February 11, 2026

VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 423864 AST-N406 423869 AST-N407 423870 AST-N409 423925 AST-N412 423936 AST-N413 423928 AST-N414 423933 AST-N415 423934 AST-N416 423935 AST-N417 423880 AST-N419

Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aerugi…

Recall #
Z-1236-2026
Affected scope
2,668,954 total
Initiated
January 5, 2026
Verify with FDA Devices →
View my saved recalls

The recall

Biomerieux Inc issued this moderate-severity FDA Devices recall — Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of….

Moderate
severity level
Class II
classification
February 11, 2026
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1236-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1236-2026) was formally reported on February 11, 2026, with the manufacturer initiating the action on January 5, 2026. It is classified under Moderate severity (Class II), with a current status of Ongoing. Biomerieux Inc is listed as the recalling firm, operating out of Hazelwood, MO. Federal records list the affected scope as 2,668,954 total.

The documented reason for this recall is: Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic form… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category — 40,409 medical devices recalls on record

-1,00001,0002,0003,0004,0005,000 20052008201120142017202020232026 1,243

Where this recall sits in the database

Severity2366872097High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified moderate severity.

Severity

Moderate

Affected scope

2,668,954 total

Related Recalls

6

6 from same agency

Product description

VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 423864 AST-N406 423869 AST-N407 423870 AST-N409 423925 AST-N412 423936 AST-N413 423928 AST-N414 423933 AST-N415 423934 AST-N416 423935 AST-N417 423880 AST-N419 423948 AST-N421 424055 AST-N428 424320 AST-N429 424321 AST-N430 424322 AST-N433 424389 AST-N439 424501 AST-N440 424502 AST-N443 424541 AST-N444 424587 AST-N467 424857 AST-N476 424934 AST-N478 425052 AST-N480 425084 AST-N481 425085 AST-N485 425182 AST-N496 425265 AST-N800 423310 AST-N802 423706 AST-N804 424634 AST-N808 424711 AST-N810 424712 AST-N811 424713 AST-N813 424722 AST-N815 425054 AST-XN09 423425 AST-XN15 423829 AST-XN17 423673 AST-XN18 423874 AST-XN20 423947 AST-XN21 424197 AST-XN22 424199 AST-XN23 424198 AST-XN24 424351 AST-XN26 424500 AST-XN28 424586 AST-XN29 424604 AST-XN30 424639 AST-XN31 424640 AST-XN32 424678 AST-XN33 424723 AST-XN34 424752 AST-XN35 424810 AST-XN36 424860 AST-XN37 424888 AST-XN38 425074 AST-XN39 425086 AST-XN40 425095 AST-XN41 425204 AST-XN42 425309 ***Update 4/2/26 -- Additional product references not manufactured at time of initial notification. These products will not be distributed within the United States.*** AST-N484 Ref. 425181 AST-N486 Ref. 425183 AST-N491 Ref. 425206 AST-N494 Ref. 425224 AST-N485 Ref. 425182 AST-XN41 Ref. 425204 AST-XN42 Ref. 425309 AST-N504 Ref. 425491

Reason for recall

Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1236-2026
Date reported February 11, 2026
Date initiated January 5, 2026
Recalling firm Biomerieux Inc
Firm location Hazelwood, MO
Affected scope 2,668,954 total
Distribution Worldwide - US Nationwide distribution.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1236-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting — recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 423864 AST-N406 423869 AST-N407 423870 AST-N409 423925 AST-N412 423936 AST-N413 423928 AST-N414 423933 AST-N415 423934 AST-N416 423935 AST-N417 423880 AST-N419 423948 AST-N421 424055 AST-N428 424320 AST-N429 424321 AST-N430 424322 AST-N433 424389 AST-N439 424501 AST-N440 424502 AST-N443 424541 AST-N444 424587 AST-N467 424857 AST-N476 424934 AST-N478 425052 AST-N480 425084 AST-N481 425085 AST-N485 425182 AST-N496 425265 AST-N800 423310 AST-N802 423706 AST-N804 424634 AST-N808 424711 AST-N810 424712 AST-N811 424713 AST-N813 424722 AST-N815 425054 AST-XN09 423425 AST-XN15 423829 AST-XN17 423673 AST-XN18 423874 AST-XN20 423947 AST-XN21 424197 AST-XN22 424199 AST-XN23 424198 AST-XN24 424351 AST-XN26 424500 AST-XN28 424586 AST-XN29 424604 AST-XN30 424639 AST-XN31 424640 AST-XN32 424678 AST-XN33 424723 AST-XN34 424752 AST-XN35 424810 AST-XN36 424860 AST-XN37 424888 AST-XN38 425074 AST-XN39 425086 AST-XN40 425095 AST-XN41 425204 AST-XN42 425309 ***Update 4/2/26 -- Additional product references not manufactured at time of initial notification. These products will not be distributed within the United States.*** AST-N484 Ref. 425181 AST-N486 Ref. 425183 AST-N491 Ref. 425206 AST-N494 Ref. 425224 AST-N485 Ref. 425182 AST-XN41 Ref. 425204 AST-XN42 Ref. 425309 AST-N504 Ref. 425491. Recalled by Biomerieux Inc. Units affected: 2,668,954 total.
Why was this product recalled?
Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 11, 2026. Severity: Moderate. Recall number: Z-1236-2026.
Where was the recalled product distributed?
Distribution: Worldwide - US Nationwide distribution..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1236-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →

Data Sources

Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported February 11, 2026.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.