FDA Devices recall · Reported March 25, 2015
It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective …
Elekta, Inc. recalled Elekta Precise Treatment Table - a moderate-severity action.
Elekta Precise Treatment Table was recalled by Elekta, Inc. in March 25, 2015. Reason: It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if…. Check the official notice for the remedy. Verify recall #Z-1238-2015 with the FDA Devices before acting.
The recall
Elekta, Inc. issued this moderate-severity FDA Devices recall-It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1238-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1238-2015) was formally reported on March 25, 2015, with the manufacturer initiating the action on January 29, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Elekta, Inc. is listed as the recalling firm, operating out of Atlanta, GA. Federal records list the affected scope as 529 US 2951 Worldwide.
The documented reason for this recall is: It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may o… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) Distribution and to the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Ch…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
529 US 2951 Worldwide
Related Recalls
6
3 from same agency
Elekta Precise Treatment Table
It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1238-2015 |
| Date reported | March 25, 2015 |
| Date initiated | January 29, 2015 |
| Recalling firm | Elekta, Inc. |
| Firm location | Atlanta, GA |
| Affected scope | 529 US 2951 Worldwide |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution and to the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 25, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.