Wallace Sure-Pro¿ Two Stage Embryo Replacement Catheter for difficult transfer with Stylet PPS623 Sure-Pro¿ Embryo Replacement Catheters are used for the introduction of embryos into the uterine cavity following in vitro fertilization. Sure-Pro Ultra¿ Embryo Replacement Catheters have the added benefit of being visible under ultrasound. These products can be used in either a one- stage or two-stage procedure. In the one-stage procedure, the clinician pushes the inner catheter through the out

Reported: May 15, 2013 Initiated: March 28, 2013 #Z-1243-2013

Product Description

Wallace Sure-Pro¿ Two Stage Embryo Replacement Catheter for difficult transfer with Stylet PPS623 Sure-Pro¿ Embryo Replacement Catheters are used for the introduction of embryos into the uterine cavity following in vitro fertilization. Sure-Pro Ultra¿ Embryo Replacement Catheters have the added benefit of being visible under ultrasound. These products can be used in either a one- stage or two-stage procedure. In the one-stage procedure, the clinician pushes the inner catheter through the outer sheath and loads the embryo prior to performing the embryo transfer into the uterus. In the two-stage procedure, the clinician places the outer sheath through the cervical canal into the uterus, then loads the embryo in the inner catheter, and advances the inner catheter through the outer sheath to perform the embryo transfer into the uterus.

Reason for Recall

Unable to pass catheter through outer sheath during embryo transfer

Details

Recalling Firm
Smiths Medical ASD, Inc.
Units Affected
4910 units
Distribution
Worldwide distribution: USA (Nationwide) and countries of: Arab Emirates (AE) Ireland (IE), Albania, Israel (IL), Argentina (AR),Italy (IT) Armenia (AM) , Japan (JP), Austria (AT) Jordan (JO), Bahrain (BH) Kuwait (KW) Belgium (BE), Lebanon (LB), Belarus , Netherlands (NL), Brazil (BR),Nigeria (NG), Bulgaria (BG) , Pakistan (PK), Canada (CA) Poland (PL), China (CN) Romania (RO), Columbia (CO), Russia (RU), Cyprus(CY) , Saudi Arabia (SA) Czech (CZ) , Singapore (SG), Denmark (OK), South Korea (KR), Egypt (EG), Spain (ES), Finland (FI), Sri Lanka (LK),Germany (DE) Turkey (TR) Ghana (GH) , Uganda (UG), Great Britain (GB) Ukraine (UA), Greece (GR) Uruguay (UY), and Hong Kong (HK).
Location
Keene, NH

Frequently Asked Questions

What product was recalled?
Wallace Sure-Pro¿ Two Stage Embryo Replacement Catheter for difficult transfer with Stylet PPS623 Sure-Pro¿ Embryo Replacement Catheters are used for the introduction of embryos into the uterine cavity following in vitro fertilization. Sure-Pro Ultra¿ Embryo Replacement Catheters have the added benefit of being visible under ultrasound. These products can be used in either a one- stage or two-stage procedure. In the one-stage procedure, the clinician pushes the inner catheter through the outer sheath and loads the embryo prior to performing the embryo transfer into the uterus. In the two-stage procedure, the clinician places the outer sheath through the cervical canal into the uterus, then loads the embryo in the inner catheter, and advances the inner catheter through the outer sheath to perform the embryo transfer into the uterus.. Recalled by Smiths Medical ASD, Inc.. Units affected: 4910 units.
Why was this product recalled?
Unable to pass catheter through outer sheath during embryo transfer
Which agency issued this recall?
This recall was issued by the FDA Devices on May 15, 2013. Severity: Moderate. Recall number: Z-1243-2013.