PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported March 8, 2017

Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.

Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with …

Recall #
Z-1252-2017
Affected scope
1,627 units (769 URF-V2 and 858 URF-V2R)
Initiated
December 12, 2016
Verify with FDA Devices →
View my saved recalls

Olympus Corporation of the Americas recalled Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R en… - a moderate-severity action.

Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R en… was recalled by Olympus Corporation of the Americas in March 8, 2017. Reason: Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section duri…. Check the official notice for the remedy. Verify recall #Z-1252-2017 with the FDA Devices before acting.

The recall

Olympus Corporation of the Americas issued this moderate-severity FDA Devices recall-Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section duri….

Moderate
severity level
2K units
affected scope
Class II
classification
March 8, 2017
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1252-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.

Reason for recall

Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1252-2017
Date reported March 8, 2017
Date initiated December 12, 2016
Recalling firm Olympus Corporation of the Americas
Firm location Center Valley, PA
Affected scope 1,627 units (769 URF-V2 and 858 URF-V2R)
Distribution Worldwide distribution-US Nationwide and country of: Canada.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1,627 units (769 URF-V2 and 858 URF-V2R) units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1252-2017) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.. Recalled by Olympus Corporation of the Americas. Units affected: 1,627 units (769 URF-V2 and 858 URF-V2R).
Why was this product recalled?
Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 8, 2017. Severity: Moderate. Recall number: Z-1252-2017.
Where was the recalled product distributed?
Distribution: Worldwide distribution-US Nationwide and country of: Canada..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported March 8, 2017. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026