Severity
Moderate
FDA Devices recall · Reported March 18, 2015
Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable ra…
Siemens Healthcare Diagnostics Inc recalled RAPIDPoint 500 Blood Gas Analyzer; Siemens Material Number: 10492730, 10696855, 10696857… - a moderate-severity action.
RAPIDPoint 500 Blood Gas Analyzer; Siemens Material Number: 10492730, 10696855, 10696857… was recalled by Siemens Healthcare Diagnostics Inc in March 18, 2015. Reason: Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater t…. Check the official notice for the remedy. Verify recall #Z-1253-2015 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics Inc issued this moderate-severity FDA Devices recall-Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater t….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1253-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1253-2015) was formally reported on March 18, 2015, with the manufacturer initiating the action on December 30, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Healthcare Diagnostics Inc is listed as the recalling firm, operating out of Norwood, MA. Federal records list the affected scope as 4376 devices.
The documented reason for this recall is: Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl. Distribution data in the federal record shows the product reached: Worldwide Distribution - US nationwide including Puerto Rico, Canada, Mexico, Albania, Algeria, Argentina, Austria, Bahrain, Belgium, Bosnia Herzegovina, Botswana, Brazil, Bulgaria, Chile, Colombia, Croatia, Czech Repub…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4376 devices
Related Recalls
6
3 from same agency
RAPIDPoint 500 Blood Gas Analyzer; Siemens Material Number: 10492730, 10696855, 10696857, 10697306. The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili.
Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1253-2015 |
| Date reported | March 18, 2015 |
| Date initiated | December 30, 2014 |
| Recalling firm | Siemens Healthcare Diagnostics Inc |
| Firm location | Norwood, MA |
| Affected scope | 4376 devices |
| Distribution | Worldwide Distribution - US nationwide including Puerto Rico, Canada, Mexico, Albania, Algeria, Argentina, Austria, Bahrain, Belgium, Bosnia Herzegovina, Botswana, Brazil, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 18, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.