FDA Devices recall · Reported March 18, 2015

VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemistry System, VITROS 3600 Immunodiagnostics System, VITROS 4600 Chemistry System, VITROS 5600 Integrated System. This product is an accessory for use with VITROS Analyzer Systems.

In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This i…

Recall #
Z-1263-2015
Affected scope
Domestic: 62 units; Foreign: 10 units
Initiated
December 29, 2014
Verify with FDA Devices →
View my saved recalls

Ortho-Clinical Diagnostics recalled VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITR… - a moderate-severity action.

VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITR… was recalled by Ortho-Clinical Diagnostics in March 18, 2015. Reason: In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanne…. Check the official notice for the remedy. Verify recall #Z-1263-2015 with the FDA Devices before acting.

The recall

Ortho-Clinical Diagnostics issued this moderate-severity FDA Devices recall-In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanne….

Moderate
severity level
Class II
classification
March 18, 2015
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1263-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1263-2015) was formally reported on March 18, 2015, with the manufacturer initiating the action on December 29, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Ortho-Clinical Diagnostics is listed as the recalling firm, operating out of Rochester, NY. Federal records list the affected scope as Domestic: 62 units; Foreign: 10 units.

The documented reason for this recall is: In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Pr… Distribution data in the federal record shows the product reached: Worldwide Distribution-US (nationwide) and the countries of Canada, Australia and England.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,861 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,787
Severity1902064468High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemistry System, VITROS 3600 Immunodiagnostics System, VITROS 4600 Chemistry System, VITROS 5600 Integrated System. This product is an accessory for use with VITROS Analyzer Systems.

Reason for recall

In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1263-2015
Date reported March 18, 2015
Date initiated December 29, 2014
Recalling firm Ortho-Clinical Diagnostics
Firm location Rochester, NY
Affected scope Domestic: 62 units; Foreign: 10 units
Distribution Worldwide Distribution-US (nationwide) and the countries of Canada, Australia and England.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1263-2015) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemistry System, VITROS 3600 Immunodiagnostics System, VITROS 4600 Chemistry System, VITROS 5600 Integrated System. This product is an accessory for use with VITROS Analyzer Systems.. Recalled by Ortho-Clinical Diagnostics. Units affected: Domestic: 62 units; Foreign: 10 units.
Why was this product recalled?
In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 18, 2015. Severity: Moderate. Recall number: Z-1263-2015.
Where was the recalled product distributed?
Distribution: Worldwide Distribution-US (nationwide) and the countries of Canada, Australia and England..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… → · All Medical Devices recalls

This recall: FDA Devices, reported March 18, 2015. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026