i-STAT CG4+ cartridge (blue), List No. 03P85-50. UDI (01) 00054749002269 - Product Usage: The i-STAT CG4+ cartridge is intended for the quantitative measurement of lactate, pH, pCO2, and pO2 in venous, arterial or capillary whole blood.

Reported: February 26, 2020 Initiated: January 15, 2020 #Z-1263-2020

Product Description

i-STAT CG4+ cartridge (blue), List No. 03P85-50. UDI (01) 00054749002269 - Product Usage: The i-STAT CG4+ cartridge is intended for the quantitative measurement of lactate, pH, pCO2, and pO2 in venous, arterial or capillary whole blood.

Reason for Recall

The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the facilities that perform these types of tests. The i-STAT CHEM8+ cartridge contains nine measured assays (sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine and hematocrit) and is categorized as a CLIA waived test for venous whole blood samples, meaning this cartridge could be used in a facility with a CLIA Certificate of Waiver as a waived test.

Details

Recalling Firm
Abbott Point Of Care Inc.
Units Affected
99,850
Distribution
US Nationwide distribution.
Location
Princeton, NJ

Frequently Asked Questions

What product was recalled?
i-STAT CG4+ cartridge (blue), List No. 03P85-50. UDI (01) 00054749002269 - Product Usage: The i-STAT CG4+ cartridge is intended for the quantitative measurement of lactate, pH, pCO2, and pO2 in venous, arterial or capillary whole blood.. Recalled by Abbott Point Of Care Inc.. Units affected: 99,850.
Why was this product recalled?
The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the facilities that perform these types of tests. The i-STAT CHEM8+ cartridge contains nine measured assays (sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine and hematocrit) and is categorized as a CLIA waived test for venous whole blood samples, meaning this cartridge could be used in a facility with a CLIA Certificate of Waiver as a waived test.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 26, 2020. Severity: Moderate. Recall number: Z-1263-2020.