Severity
Moderate
FDA Devices recall · Reported September 6, 2023
Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results
Siemens Healthcare Diagnostics, Inc. recalled Atellica CH Total Protein II- In vitro diagnostic use in the quantitative determination o… - a moderate-severity action.
Atellica CH Total Protein II- In vitro diagnostic use in the quantitative determination o… was recalled by Siemens Healthcare Diagnostics, Inc. in September 6, 2023. Reason: Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Trigly…. Check the official notice for the remedy. Verify recall #Z-1269-2023 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Trigly….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1269-2023 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1269-2023) was formally reported on September 6, 2023, with the manufacturer initiating the action on January 27, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. Siemens Healthcare Diagnostics, Inc. is listed as the recalling firm, operating out of Tarrytown, NY. Federal records list the affected scope as 13071 units.
The documented reason for this recall is: Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results Distribution data in the federal record shows the product reached: Nationwide Foreign: Country Albania Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Finland Fr…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
13071 units
Related Recalls
6
3 from same agency
Atellica CH Total Protein II- In vitro diagnostic use in the quantitative determination of total protein in human serum and plasma (lithium heparin) using the Atellica CH Analyzer. SMN: 11097604
Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1269-2023 |
| Date reported | September 6, 2023 |
| Date initiated | January 27, 2023 |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Firm location | Tarrytown, NY |
| Affected scope | 13071 units |
| Distribution | Nationwide Foreign: Country Albania Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Finland France Germany Greece … |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 6, 2023. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.