PlainRecalls
FDA Devices Verify with FDA Devices → Moderate Class II Terminated

Estradiol II Elecsys and cobas e analyzers REF 03000079 190; M 6.5mL, R1 8 mL, R2 8 mL, IVD, For USA: CONTENT M Streptavidin-coated microparticles 0.72 mg/mL; R1 Biotinylated polyclonal anti-estradiol antibody (rabbit) 45 ng/mL; Mesterolone 130 ng/mL; R2 Estradiol derivative, labeled with ruthenium complex 2.75 ng/mL Product Usage: Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma. The electrochemiluminescence immunoassay ECLIA is intended

Reported: April 6, 2016 Initiated: March 7, 2016 #Z-1277-2016 102,312 units

Roche Diagnostics Operations, Inc. issued this FDA Devices recall on April 6, 2016. Classified as Moderate severity (Class II). Approximately 102,312 units are affected. The recall was issued because: Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitori…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-1277-2016) was formally reported on April 6, 2016, with the manufacturer initiating the action on March 7, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Operations, Inc. is listed as the recalling firm, operating out of Indianapolis, IN. Federal records indicate 102,312 units are affected, a scale large enough to require multi-state distribution tracking.

The documented reason for this recall is: Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitoring patient being treated with fulvestrant. Customers should be informed that fulvestrant will incre… Distribution data in the federal record shows the product reached: US Nationwide Distribution including Puerto Rico.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 10 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

102,312

Related Recalls

6

6 from same agency

Product Description

Estradiol II Elecsys and cobas e analyzers REF 03000079 190; M 6.5mL, R1 8 mL, R2 8 mL, IVD, For USA: CONTENT M Streptavidin-coated microparticles 0.72 mg/mL; R1 Biotinylated polyclonal anti-estradiol antibody (rabbit) 45 ng/mL; Mesterolone 130 ng/mL; R2 Estradiol derivative, labeled with ruthenium complex 2.75 ng/mL Product Usage: Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.

Reason for Recall

Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitoring patient being treated with fulvestrant. Customers should be informed that fulvestrant will increase the apparent concentration of estradiol in women being treated with this drug. If treatment with fulvestrant has been altered or discontinued as a result of falsely elevated e

Details

Units Affected
102,312
Distribution
US Nationwide Distribution including Puerto Rico.
Location
Indianapolis, IN

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1277-2016
Date reported April 6, 2016
Date initiated March 7, 2016
Recalling firm Roche Diagnostics Operations, Inc.
Units affected 102,312
Distribution US Nationwide Distribution including Puerto Rico.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

102,312 units affected — multi-state distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Estradiol II Elecsys and cobas e analyzers REF 03000079 190; M 6.5mL, R1 8 mL, R2 8 mL, IVD, For USA: CONTENT M Streptavidin-coated microparticles 0.72 mg/mL; R1 Biotinylated polyclonal anti-estradiol antibody (rabbit) 45 ng/mL; Mesterolone 130 ng/mL; R2 Estradiol derivative, labeled with ruthenium complex 2.75 ng/mL Product Usage: Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.. Recalled by Roche Diagnostics Operations, Inc.. Units affected: 102,312.
Why was this product recalled?
Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitoring patient being treated with fulvestrant. Customers should be informed that fulvestrant will increase the apparent concentration of estradiol in women being treated with this drug. If treatment with fulvestrant has been altered or discontinued as a result of falsely elevated e
Which agency issued this recall?
This recall was issued by the FDA Devices on April 6, 2016. Severity: Moderate. Recall number: Z-1277-2016.
Where was the recalled product distributed?
Distribution: US Nationwide Distribution including Puerto Rico..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1277-2016) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).