FDA Devices recall · Reported May 29, 2019

NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system w…

Recall #
Z-1279-2019
Affected scope
37
Initiated
October 22, 2018
Verify with FDA Devices →
View my saved recalls

Neusoft Medical Systems Co., Ltd. recalled NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner Syst… - a moderate-severity action.

NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner Syst… was recalled by Neusoft Medical Systems Co., Ltd. in May 29, 2019. Reason: Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during …. Check the official notice for the remedy. Verify recall #Z-1279-2019 with the FDA Devices before acting.

The recall

Neusoft Medical Systems Co., Ltd. issued this moderate-severity FDA Devices recall-Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during ….

Moderate
severity level
37 units
affected scope
Class II
classification
May 29, 2019
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1279-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Reason for recall

Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1279-2019
Date reported May 29, 2019
Date initiated October 22, 2018
Recalling firm Neusoft Medical Systems Co., Ltd.
Firm location Shenyang, N/A
Affected scope 37
Distribution US Nationwide Distribution.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

37 units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1279-2019) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled? ▼
NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.. Recalled by Neusoft Medical Systems Co., Ltd.. Units affected: 37.
Why was this product recalled? ▼
Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 29, 2019. Severity: Moderate. Recall number: Z-1279-2019.
Where was the recalled product distributed? ▼
Distribution: US Nationwide Distribution..
Should I stop using a recalled medication or medical device? ▼
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported May 29, 2019. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026