ModerateClass IITerminated

FDA Devices recall · Reported April 6, 2016

Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 AlluraXperFDlO C, 722002 Allura Xper FD1O F, 722003 Allura Xper FD1O, 722005 Allura Xper FD1O/10, 722006 Allura Xper FD2O, 722008 Allura Xper FD2O Biplane, 722010 Allura Xper FD1O, 722011 Allura Xper FD1O/10, 722012 Allura Xper FD2O, 722013 Allura XPER FD2O BIPLANE, 722014 Allura Xper FD1O OR Table, 722015 Allura Xper FD2O OR Table, 722019 Allura Xper FD1O/10 OR Table, 722020 Allura Xper FD2O Biplane OR Table, 722

Loss of key image functionality due to a bent pedal of the Footswitch.

Recall #
Z-1280-2016
Affected scope
4180 devices
Initiated
September 24, 2014
Verify with FDA Devices →
View my saved recalls

Philips Medical Systems, Inc. recalled Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 Allura… — a moderate-severity action.

Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 Allura… was recalled by Philips Medical Systems, Inc. in April 6, 2016. Reason: Loss of key image functionality due to a bent pedal of the Footswitch.. Check the official notice for the remedy. Verify recall #Z-1280-2016 with the FDA Devices before acting.

The recall

Philips Medical Systems, Inc. issued this moderate-severity FDA Devices recall — Loss of key image functionality due to a bent pedal of the Footswitch..

Moderate
severity level
4K units
affected scope
Class II
classification
April 6, 2016
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1280-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1280-2016) was formally reported on April 6, 2016, with the manufacturer initiating the action on September 24, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Medical Systems, Inc. is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 4180 devices.

The documented reason for this recall is: Loss of key image functionality due to a bent pedal of the Footswitch. Distribution data in the federal record shows the product reached: Worldwide Distribution - US (nationwide) and Internationally to: CZECH REPUBLIC ALBANIA ALGERIA ARGENTINA AUSTRALIA AUSTRIA AZARBAIJAN BAHAMAS BANGLADESH BELGIUM BRAZIL BULGARIA CAYMAN ISLANDS CHILE CHINA…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category — 40,409 medical devices recalls on record

-1,00001,0002,0003,0004,0005,000 20052008201120142017202020232026 1,243

Where this recall sits in the database

Severity2366872097High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger — we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

4180 devices

Related Recalls

6

6 from same agency

Product description

Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 AlluraXperFDlO C, 722002 Allura Xper FD1O F, 722003 Allura Xper FD1O, 722005 Allura Xper FD1O/10, 722006 Allura Xper FD2O, 722008 Allura Xper FD2O Biplane, 722010 Allura Xper FD1O, 722011 Allura Xper FD1O/10, 722012 Allura Xper FD2O, 722013 Allura XPER FD2O BIPLANE, 722014 Allura Xper FD1O OR Table, 722015 Allura Xper FD2O OR Table, 722019 Allura Xper FD1O/10 OR Table, 722020 Allura Xper FD2O Biplane OR Table, 722022 Allura Xper FD1O OR Table, 722023 Allura Xper FD2O OR Table, 722024 Allura Xper FD1O/10 OR Table, 722025 Allura Xper FD2O Biplane OR Table, 722026 Allura Xper FD1O, 722027 Allura Xper FD1O/10, 722028 Allura Xper FD2O, 722029 Allura Xper FD2O/10, 722031 Allura CV2O, 722033 Allura Xper FD1O OR Table, 722034 Allura Xper FD1O/10 OR Table, 722035 Allura Xper FD2O OR Table, 722036 Allura Xper FD2O/10 OR Table, 722038 Allura Xper FD2O/20, 722039 Allura Xper FD2O/20 OR Table, 722058 Allura Xper FD20115, 722059 Allura Xper FD2O/15 OR Table, 722123 Field extensions Xper cardio systems, 722124 Field extensions Xper vascular systems, 722126 SmartPath to AlluraClarity cardio, 722127 SmartPath to AlluraClarity vascular, 722133 Field ext. Xper cardio systems R7.6, 722134 Field ext. Xper vascular systems R7.6, 722400 Cardio Vascular-Allura Centron, 889006 Diamond Select Allura Xper FD1 0 The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Tis included, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper FD series is compatible with a hybrid Operating Room. FD10 is compatible with specific magnetic navigation systems

Reason for recall

Loss of key image functionality due to a bent pedal of the Footswitch.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-1280-2016
Date reported April 6, 2016
Date initiated September 24, 2014
Recalling firm Philips Medical Systems, Inc.
Firm location Andover, MA
Affected scope 4180 devices
Distribution Worldwide Distribution - US (nationwide) and Internationally to: CZECH REPUBLIC ALBANIA ALGERIA ARGENTINA AUSTRALIA AUSTRIA AZARBAIJAN BAHAMAS BANGLADESH BELGIUM BRAZIL BULGARIA CAYMAN ISLANDS CHILE CHINA COLUMBIA COSTA RI…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

4180 devices units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1280-2016) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting — recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Foot Switches used with the following systems: Philips Allura Xper Systems; 722001 AlluraXperFDlO C, 722002 Allura Xper FD1O F, 722003 Allura Xper FD1O, 722005 Allura Xper FD1O/10, 722006 Allura Xper FD2O, 722008 Allura Xper FD2O Biplane, 722010 Allura Xper FD1O, 722011 Allura Xper FD1O/10, 722012 Allura Xper FD2O, 722013 Allura XPER FD2O BIPLANE, 722014 Allura Xper FD1O OR Table, 722015 Allura Xper FD2O OR Table, 722019 Allura Xper FD1O/10 OR Table, 722020 Allura Xper FD2O Biplane OR Table, 722022 Allura Xper FD1O OR Table, 722023 Allura Xper FD2O OR Table, 722024 Allura Xper FD1O/10 OR Table, 722025 Allura Xper FD2O Biplane OR Table, 722026 Allura Xper FD1O, 722027 Allura Xper FD1O/10, 722028 Allura Xper FD2O, 722029 Allura Xper FD2O/10, 722031 Allura CV2O, 722033 Allura Xper FD1O OR Table, 722034 Allura Xper FD1O/10 OR Table, 722035 Allura Xper FD2O OR Table, 722036 Allura Xper FD2O/10 OR Table, 722038 Allura Xper FD2O/20, 722039 Allura Xper FD2O/20 OR Table, 722058 Allura Xper FD20115, 722059 Allura Xper FD2O/15 OR Table, 722123 Field extensions Xper cardio systems, 722124 Field extensions Xper vascular systems, 722126 SmartPath to AlluraClarity cardio, 722127 SmartPath to AlluraClarity vascular, 722133 Field ext. Xper cardio systems R7.6, 722134 Field ext. Xper vascular systems R7.6, 722400 Cardio Vascular-Allura Centron, 889006 Diamond Select Allura Xper FD1 0 The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Tis included, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper FD series is compatible with a hybrid Operating Room. FD10 is compatible with specific magnetic navigation systems. Recalled by Philips Medical Systems, Inc.. Units affected: 4180 devices.
Why was this product recalled?
Loss of key image functionality due to a bent pedal of the Footswitch.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 6, 2016. Severity: Moderate. Recall number: Z-1280-2016.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - US (nationwide) and Internationally to: CZECH REPUBLIC ALBANIA ALGERIA ARGENTINA AUSTRALIA AUSTRIA AZARBAIJAN BAHAMAS BANGLADESH BELGIUM BRAZIL BULGARIA CAYMAN ISLANDS CHILE CHINA COLUMBIA COSTA RICA CROATIA CUBA CZECHOSLOVAKIA DENMARK DOMINICAN REPUB ECUADOR EGYPT ESTONIA FINLAND FRANCE GABON GERMANY GREECE HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KENYA KOREA SOUTH KUWAIT LATVIA LEBANON LIBYA LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVIA MONGOLIA MOROCCO NEPAL NETHERLANDS NETHERLANDS ANT NEW ZEALAND NORWAY OMAN PAKISTAN PALESTINE PANAMA Papua New Guinea PARAGUAY PERU PHILIPPINES POLAND PORTUGAL PUORTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERAT SAUDI ARABIA SERBIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SUDAN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TUNISIA TURKEY UKRAINE UNITED ARAB EMI United Arab Emirates UNITED KINGDOM URUGUAY USA UZBEKISTAN VENEZUELA VIETNAM YEMEN GUADELOUPE SYRIAN ARAB REPUBLIC.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1280-2016) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →

Data Sources

Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported April 6, 2016.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.